SCS-818 4-pole butterworth SCS-819 4-pole bessel

get the pdf data sheet
Click here to get prices

8-Channel Amplifier and Low-Pass Filter Module for the SCS-800/801 Signal Conditioning System


  • 8 amplifiers/filters per module
  • AC or DC couple the signal input
  • Individual channel independent corner frequency selector
  • Individual channel independent instrumentation amplifier control
  • gain steps of x1, x2, x5, x10, x20, x50, x100, x500, and x1000 
  • 12 selectable corner frequency control sources can be used to control any channel
  • Corner frequencies can be set to any frequency in the filter bandwidth all under software control
  • Butterworth bandwidth from 0.5Hz to 70kHz on SCS-818
  • Bessel bandwidth from  0.5Hz to 70kHz on SCS-819
  • Programmable A/D synchronization clock output
  • ±10V input and output
  • Automatic DC offset compensation removes the need for a software calibration procedure
  • Offset guaranteed <1mV for all gain settings
  • Differential output, proprietary DC transformer reduces common mode noise problems to both single-ended and differential A/D hosts
  • Onboard channel and module multiplexers
  • Input cable shield drive reduces common mode problems
  • Common mode voltage protection to ±40V
  • Optional mono to bi polar offset switch increases signal resolution at the A/D.
  • Non-volatile memory retains all settings during power interruptions 


Overview

The SCS-818 and SCS-819 amplifier/filter module provides eight 4-pole low-pass filters that can be individually configured for a cutoff-frequency from 0.5Hz to 70kHz(SCS-818) and 0.5Hz to 70kHz(SCS-819). Each of the SCS-818/SCS-819's eight channels features an instrumentation amplifier with software-selectable gains of 1 to 1000 in 1-2-5 steps.

Input

The SCS-818/SCS-819 has eight separate differential inputs, which may be DC or AC coupled to the instrumentation amplifier. The amplifier has an input voltage range of ±10V and is protected to ±40V. The instrumentation amplifier plus a second programmable amplifier provide gain selection from 1 to 1000. Offset is factory calibrated below the rated specification at both input and output stages of the instrumentation amplifier.

Filter

The output of the gain stage is followed by separate low-pass filter stages. One-pole filters reduce clock feed through. Each channel's -3 dB corner frequency is independently software controlled, selectable from any of eight internal or four external sources.

Automatic offset compensation nulls errors due to drift and aging.

Output

A proprietary DC transformer circuit provides true differential outputs, with very low 10 ohm impedance. This circuit guarantees that common-mode noise will not effect the signal at the receiving end. This circuit also allows for use of single ended inputs on the A/D converter, effectively doubling the number of A/D channels without sacrificing common-mode rejection. Outputs are scaled to provide ±10V. Each channel output is available as a differential signal on the output connector and can be wired to the data acquisition receiver as either a single-ended or differential input.

The SCS-800/801 system provides a variety of methods for bringing outputs to the data acquisition host:

  • All output signals can be wired in parallel to the host for simultaneous sampling.
  • All output signals can be time-multiplexed onto one set of analog output lines via the SCS-818/SCS-819's front-panel output connector.
  • All output signals can be time-multiplexed with all other module outputs via the SCS-800 back plane to one connector on the SCS-804 System Controller Module. This method provides channel multiplexing, or channel and module multiplexing.

The connection methods are independent. Some modules may be multiplexed onto the back plane, and others multiplexed out their own front panel, while other modules can have all channels as independent outputs.

CE mark


Specifications

Coupling
AC cutoff 0.03Hz typical
Input
Voltage range ±10V
Common mode rejection 75dB min, 86dB typ (gain = 1)
Noise (RTI @G=1) 30nV per Hz
Coupling AC or DC
Impedance 2Mohm/20pF
Protection ±40V
Bias current ±2pA typ, ±100pA max
Offset
Voltage ±0.1mV typ, ±1.0mV max
Current ±1pA typ, ±100pA max
Gain
Steps 1,2,5,10,20,50,100,200, 500,1000
Accuracy ±0.08dB max
Output
Impedance 10ohms
Voltage range ±10V
General
Temperature, operating 0°C to 55°C
Temperature, storage -25°C to 85°C
Humidity 0% to 95% non-condensing
Multiplexer
Minimum switching time 100µS
Settling time (controller card specification) 10µS to 0.1%
Connector
Input and output Amp 787096-1
Latch Amp 787003-3
Mating Connector
Input and output Amp 787131-1
Backshell Amp 787133-1

PHONE 949-515-1400        FAX 949-515-4724       E-MAIL productinfo@alligatortech.com

 